CO2 Incubator
MAKE DINESH SCIENTIFIC MODEL DS-CO2I-170L
PRODUCT DESCRIPTION
With the support of our adroit professionals, we provide our clients a high quality CO2 Incubator. Our provided incubator is made using high grade components, procured from reliable sources of the industry in compliance with international standards under the guidance of skilled professionals. Owing to its unique feature, it is widely acclaimed and appreciated in the market.
SAILENT FEATURES:
- Superior strength
- Robust design
- Impeccable finish
- Micro-computer chip – PID Control System. …
- Advance Ultrasonic Sensors. …
- High-Precision Platinum Resistance Temperature Sensor. …
- Air Circulation. …
- Auto-Start Function. …
- Solenoid Valve. …
- Temperature Controller & Overheating Relay Function.
FUNCTIONS OF CO2 INCUBATOR:
- The purpose of a CO2 incubator is to maintain an optimal environment for cell growth, by providing carbon dioxide control in a humidified atmosphere with constant temperature.
- chamber capacity, the volume of the chamber that is available for allowing growth; temperature range, the range of temperatures the incubator can control; temperature control accuracy, the closeness of an indication or reading of a measurement device to the actual value of the quantity being measured.
SCOPE OF APPLICATIONS:
- The CO2 incubator is used for providing and maintaining the optimum temperature, controlling the environment’s humidity, and regulating the levels of necessary gases such as oxygen and carbon dioxide. Aside from that, the incubator is also responsible for protecting the precious cultures from microbial contamination.
- Incubators are used to grow microbial culture or cell cultures. Incubators can also be used to maintain the culture of organisms to be used later. Some incubators are used to increase the growth rate of organisms, having a prolonged growth rate in the natural environment.
TECHNICAL DATA:
PARAMETER | VALUE |
Capacity | 170 L |
Temperature management | 4°C above ambient to 50°C |
Control increment | 0.1 °C |
Temperature accuracy | + 0.4 °C at 37°C |
Points in the chamber | 27 points in the chamber |
Ambient | 22°C |
Temperature stability | + 0.1°C at 37°C and ambient 22°C, |
Temperature uniformity | + 0.3°C at 37°C and ambient 22°C |
CO2 gas range | 0.1 – 20% |
Control increment | 0.1%, |
Accuracy | + 0.3% |
Specified Relative Humidity (RH) | 37°C and ambient 22°C |
Gas uniformity | + 0.1% at 37°C and ambient 22°C across the chamber |
CO2 recovery rate | 6 min after door opening and closing event to attain 5% CO2 |
Optional High-Temperature Disinfection [HTD] | 140 °C for 2 hours. Entire HTD cycle [including the time for warming up and cooling down to incubation temperature (37oC)] will not take more than 14 hours |
BMS Relay | Will have BMS relays built in and option to incorporate onto Data monitoring and documentations modules. |
Door | The door hinges, associated cable and other accessories will be robust and stringently tested. |
Display | Large backlit display for control of temperature and alarms |
Inner Door | Separate single inner glass door for monitoring of samples without disturbing conditions of the chamber. |
Gas Comsumption | An inline pressure regulator to ensure less gas consumption and prevent overshooting of pressure which shortens life span of incubator. |
Inner Chamber Material | Single stainless-steel sheet with deep-drawn and seamless design with no corners, welds or joints for higher capacity and ease of cleaning |
Heatring element | Six-sided direct heating elements to ensure even distribution of heat throughout the entire incubator chamber. |
Humidity Tray | A removable humidity tray for easy cleaning and refilling of distilled water |
Fan | Will be “fan less” design to reduce chance of contamination, reduce noise level, minimum air turbulence and bigger usable capacity. |
Sensor | NDIR type CO2 sensor with auto-calibration feature to ensure accuracy of sensor automatically and will withstand at least 180 °C during high temperature disinfection. |
Shelves | 3 perforated stainless-steel shelves. |
Access port | Access ports at the back of the chamber to allow for external probes, etc., for third party monitoring of chamber conditions. |
Machine should be supplied | CO2 gas cylinder(1 no) & 2 stage CO2 regulator(1 no) |
Reviews