BIOSAFETY CABINET
MAKE DINESH SCIENTIFIC MODEL DS-BC-4x2x2
DESCRIPTION:
A specialized piece of lab equipment called a biosafety cabinet (BSC) is made to offer a controlled atmosphere for handling hazardous compounds, especially biological agents that could endanger the worker, the surrounding area, or the samples they are handling. Maintaining the integrity of pathogen-related investigations and safeguarding researcher safety depend on the use of biosafety cabinets.
FEATURES:
A canopy exhaust connection allows the Type A2 cabinet’s HEPA-filtered exhaust air to be either vented outside or back into the space.
A fraction of mixed downflow and inflow air from the same plenum is known as HEPA-filtered downflow air.
All biologically contaminated ducts and plenums have negative pressure air surrounding them. The work area is protected from the negative pressure by double-wall plenums.
High-efficiency particulate air (HEPA) filters, which are incredibly effective, are passed via air drawn through a pre-filter. The HEPA filter’s lifespan is significantly increased by the installed prefilter. Equipped with a HEPA filter, it complies with EU EU-14 Grade regulations and has an efficiency rating of
Over 99.995% for 0.3 μ. The filter is composed of imported microfine glass fiber media that is fire retardant and water resistant. Higher air flow volumes are accommodated by the design of these filters.
Prefilters with an efficiency of down to 10 microns are made of non-woven synthetic material, complying with EU-4/G-4 Grade and featuring HDPE mesh and Al expanded mesh on the air-leaving side. It is possible to clean these filters.
Placed atop a robust tubular stand featuring retractable castors.
Aerodynamic airflow grills guard against obstructions and ensure safety.
Available with UV and fluorescent interlock
PROTECTION OF THE CASS II BIOSAFETY CABINET:
Particulate matter is dispersed throughout the chamber by high-velocity jet nozzles that introduce filtered air into it. High-velocity air is blasted through HEPA filters.
The contaminated air is drawn in by the lower part of the air showers and then passes through a pre-filter placed at the base to reach the blower supply plenum.
This is a continuous process that keeps the air inside the chamber filtered and recirculated without any interaction with the outside world.
FRONT SASH/WINDOW:
For a clear view of the interior, the front sash is made of toughened transparent glass. Furthermore, it offers the best resistance and protection against dangerous UV rays, shielding the wearer from UV radiation exposure.
VALVE-FITTED DRAIN TROUGH:
If there are any liquid spills, they are easily cleaned up and disposed of in the base trough.
UV LAMP DECONTAMINATION:
The motor blower and UV light, which emit UV rays with a wavelength of 253.7 nanometers, are interlocked, and they turn on and off in tandem with the opening and closing of the front sash. A timer (option) is used by UV lights to regulate the decontamination cycle and extend their lifespan. After the predetermined amount of time for decontamination has passed, the UV lamp turns off and turns the fluorescent lamp ON.
TECHNICAL DETAILS:
Type Class II A2 Biosafety Cabinet suited for cell culture applications.
Motor Type: Outfitted with DC motors.
Size The work chamber has a single- or multi-piece stainless steel workbench and is 1200mm (4 feet) in length. The sidewalls are composed of stainless steel (304) or glass.
Air Circulation Maintains 30% exhaust and 70% air recirculation in accordance with Class II A2 regulations. Sensors continuously track and show the airflow.
Control System Operated by a microprocessor-based control system.
Filter Type Uses HEPA (H14 HEPA EN 1822) / ULPA filters with 99.990% efficiency for 0.3-micron particle filtration.
Cabinet Construction Made from galvanized steel with a durable powder-coated finish.
UV Lamp Includes a UV lamp with an adjustable timer.
Display & Control Panel features an external digital display and control panel with an integrated hour counter.
Built-in alarms notify users of any biosafety risks.
Noise Level Operates at a noise level of 70 dBA.
Intuitive Control Panel displays downflow and inflow velocities, digital airflow verification, and alert indications to provide real-time safety and performance data.
Illumination Fitted with a white fluorescent lamp for workspace lighting.
Electrical Outlet Contains a 3-pin 5 A socket inside the chamber, compatible with Indian Plug-top.
Gas Port: Integrated universal gas valve/port within the cabinet.
Power supply supports 220-240 V / 50-60 Hz power input.