LAMINAR AIR FLOW
MAKE DINESH SCIENTIFIC
MODEL DS-LAF-4X2X2
DESCRIPTION:
Laminar airflow is the steady, unidirectional flow of air in parallel layers with little to no turbulence and a constant velocity. In many contexts, including scientific, medical, and industrial ones where preserving a sterile and clean atmosphere is essential, this regulated airflow is frequently used. Systems with laminar airflow are made to lessen the possibility of airborne particles like dust, bacteria, and other pollutants contaminating the environment.
KEY FEATURES AND COMPONENTS OF LAMINAR AIRFLOW SYSTEMS INCLUDE:
AIR FILTRATION:
Ultra-low penetration air (ULPA) -efficiency particulate air (HEPA) filters are commonly used in laminar airflow systems. By removing airborne particles, these filters guarantee that there are no
pollutants in the air entering the laminar flow area.
ONE-WAY FLOW:
Typically, the air moves from one side to the other or from the top to the bottom of the working area. By guaranteeing that particles are removed from the clean region, this helps avoid cross-contamination.
STEADY SPEED:
Laminar airflow systems keep the air velocity in the workplace constant and even. In order to reduce turbulence, the airflow system is carefully designed and calibrated.
SANITIZED WORK AREA:
Cleanrooms, labs, hospitals, and other places where keeping a sterile or regulated environment is essential frequently employ these systems. Medical procedures, pharmaceutical processing, and semiconductor manufacturing are a few examples.
CABINET CONSTRUCTION
• The cabinet (Vertical) manufactured with cold rolled steel plate with corrosion-resistant epoxy resin powder sprayed on the surface, resistant to Fumigation solution, acid, and alkali corrosion.
• Stainless steel (Grade 304) operation platform without screws and arc angle corners to prevent contamination.
• a spill-retaining design in the recessed work area of the table for accidental liquid microbial broth spill.
• Side of the LAF workbench will be covered with UV-resistant toughened/tempered glass for visibility and light.
SAFETY
• An audible alert will alert the operator if the window is raised higher than is advised. A special feature of the system will sound an alarm to cut off the air supply if the negative pressure exceeds a certain threshold, preventing contamination from entering the laboratory.
• Antimicrobial coating to increase safety by preventing bacteria development
TECHNICAL DATA:
MODEL DS-LAF-4X2X2
Design Vertical Model
Display Control panel with a digital display showing real-time airflow for better performance monitoring
Filter HEPA/ULPA filters with 99.99% efficiency for 0.1 to 0.3 microns, with integral metal guards & filter frame gaskets
Blower Motor Laminar Airflow include EMC DC motors
Light
Light Fluorescent light (>300 Lux) with low-energy choke less design to withstand voltage fluctuations, placed outside working zone to avoid turbulence
UV Lamp Located in the working zone (40 micro watts/square cm at 254 nm) and positioned to protect operator’s eyes
UV Light Programming Available with timer
Front door
Front Door Material UV-resistant toughened/tempered glass
Glass Thickness 5 mm
Door Type Vertical sliding (one-piece) with counterweight arrangement for easy sliding and fingertip control
UV Safety Mechanism Integrated with front door operation: UV automatically switches “OFF” when the sash door is opened and switches “ON” when the door is closed to prevent accidental exposure to the operator
Dimension Work area approximately 4 feet wide (W × H × D – 1200 × 600 × 600 mm) with shutter opening
Air Flow Velocity 0.40 m/s
Noise Level <65 decibels for a better working
Castor Wheels 4 castor wheels with lock for easy movement
Controlling System Cabinet control system with mini helix pressure gauge, motor voltage regulator, audible & visual window alarm, main & outlet power circuit breakers, and power switches for exterior fluorescent lights, UV lights, interior outlets, and blower motor
ACCESSORIES
• Universal service fittings for gas and air will be provided with the system.
• Switches & Electrical Sockets
• Provision for 15/5 amp sockets