BIOSAFETY CABINET
MAKE DINESH SCIENTIFIC MODEL DS-BC-4x2x2
PRODUCT DESCRIPTION
Our clientele can avail from us a wide range of Biosafety Cabinets. Our experts are using the advanced technology and sophisticated component to develop these equipments. Our presented equipments are made keeping in mind the precise needs of the customers. Biological Safety Cabinets Provide Product, Operator And Environmental Protection And Are Suitable For Work With Trace Amounts Of Toxic Chemicals And Agents Assigned To Biological Safety Levels
SALIENT FEATURES:
- Top In Work
- High Strength
- Easy To Use
- Large front panel display and user friendly control panel
- 10° slanted view screen provides highly visible and easily accessible area
- Work area is surrounded by negative pressure , double-wall plenums for superior protection
- Sturdy construction of cabinet protects from air leakage from the working zone
- Display of Life of UV lamp, HEPA filter and temperature
- Caster wheels for ease of mobility
- Levelling jack for vibration free operations
SCOPE OF APPLICATIONS:
A biological safety cabinet (BSC) is the most important equipment used in diagnostic and research laboratories in order to safeguard the product, the person, and the environment.
TECHNICAL DATA:
PARAMETERS | VALUES |
Make | DINESH SCIENTIFIC |
Model | DS-BC-4X2X2 |
Type Of Configuration | Benchtop |
Width Of Working Area (Mm) | 4ft |
Depth Of The Working Area | 2ft |
Material Used For Working Area | Stainless Steel 304 Grade |
Material Used For Cabinet Area | Stainless Steel 304 Grade |
Height Of Working Area | 2ft |
Inlow Velocity In Feet Per Min | 54 |
Intensity Of Light In Lux | 600 |
Display | LED |
Type Of Air Flow Pattern | 70% Air Recirculated Through Filter 30%Air Exhausted |
Maximum Opening (Mm) | 200 |
Filter | Hepa Lter 99.999 % Efcient |
Class | 100 |
U V Lamp Provided | With Separate Switch |
Power SUPPLY | 230v +/- 10v,50hz |
Noise Level(Db) | 65db |
Voltage | 230v+/- |
ADDITIONAL DATA:
- The Bio-safety cabinets will be Class II Type A2 4 feet width with front window will be a 10″ sash opening and coated steel interior
- The Bio Safety Cabinet will include two DC motors
- The motor will automatically adjust the airflow speed without the use of a damper to ensure continuous safe working conditions even without maintenance adjustments
- Cabinet will use a pressure sensor (rather than anemometer) to detect pressure drop across the supply filter rather than in just one point across the down flow
- The cabinet will display the inflow and down flow air velocities and will incorporate an LED Indicator visual & audible alarm to indicate excessive HEPA filters loading blower failure, airflow speed failure Incorrect window position.
- The cabinet noise level will be less than 65 dB(A) and lights on and fan will consume less than 400 watts
- The cabinet will automatically reduce fan/blower motor speed to 30% when the front window sash is in closed position to ensure reduced energy consumption when the cabinet is not is use.
- UV light will be programmable to allow for specific exposure times from 0 to 24 hours
- The Cabinet will have provision to fit taps for Vacuum, Water and Noncombustible gas